| SENSOR PROBE FORMAT |
| HPX |
-XXX |
X |
-X |
-XX |
-XX |
-XX |
Probe
Type
|
Sensor
Diameter
|
Probe
Length
(Example 1 only)
|
Operating
Temperature
|
Cable
Type
|
Cable
Length
|
Connector
Type
|
C: Cylindrical
B: Button
T: Threaded
R: Rectangular
S: Flat
D: Dual Flat |
"In Thousandths"
(.0254)
|
A: 0.500" (12.70)
B: 0.625" (15.87)
C: 0.750" (19.05)
D: 0.875" (22.23)
E: 1.000" (25.40)
F: 1.250" (31.75)
G: 1.500" (38.10)
H: 1.750" (44.45)
I: 2.000" (50.8)
|
A: 32 to 300ºF
(0 to 149ºC)
E: -100 to 400ºF
(-73 to 204ºC)
V: -100 to 1600ºF
(-73 to 871ºC)
S: -100 to 1832ºF
(-73 to 1000ºC) |
L2: .093 (2.36)
L3: .063 (1.60)
I2: .125 (3.18)
N: .125 (3.18)
ND: .165 (4.19)
T: .063 (1.60) |
In feet
(about 0.3m)
|
M: 10-32
Microdot
Male
B: BNC
Male
FX: MCX
Female
MX: MCX
Male |
Probe Thickness
(Example 2 only)
|
|
A: .093" (2.36)
B: .065" (1.65)
C: .125" (3.18)
D: .025" (0.64)
|
|
|
|
Standard sizes in bold
Please consult factory for availibility of probe selection |
|
|
|
EXAMPLE (1): HPT-150X-A-L2-5-B |
|
EXAMPLE (2): HPB-75X-E-L3-5-M |
|
|
 |
|
 |
|
|
SPECIFICATION & PROBE NOTES
Dimensions: All dimensions are shown as typical only. Please contact the factory for exact dimensions.
Maximum Range: Same as the diameter of the sensor element. (Minimum range is touch or 0.1% of full scale.)
LINEAR Range: Two thirds (0.67) times the sensor diameter typical.
Probe Interchangeability: ±10% of full scale maximum change, ±2% typical.
Probe Material Composition: A and E, stainless 303: V, Inconel 600, other materials available.
Calibrations: There are 21 numbered, standard calibrations available, please consult factory and specify when ordering.
Range:
Standard listed ranges are ratios from 0 to10.000VDC analog output, but each probe may be calibrated with several range versus sensitivity calibrations. Calibration sensitivity from 0.001VDC/0.001" (0.0254mm) to 10.000VDC/0.001(0.0254mm) are available.
Range is defined as from touch to full scale reading, ±50% of range = ±10.000VDC calibration optional.
Grounding: -D dip coatings are available to prevent the electronically driven sensor/guard elements from being accidentally earth grounded and shorted.
Extension Cables: Many extension cables are available depending upon connector types. Consult factory for proper part numbers. Example: EC-D-L2-5 is a standard 5 foot extension cable with female BNC to male BNC connectors. |
|
|
|
| CABLE SPECIFICATIONS |
Cable Temperature Range for A & E Probes, -65°F to 446°F (-54°C to 230°C)
|
| Type L2 |
OD .093 (2.36) |
Teflon |
~ 30 pF/ft (90/m) |
| Type L3 |
OD .063 (1.60) |
Teflon |
~ 26 pF/ft (78/m)
|
| Type I2 |
OD .125 (3.18) |
Teflon |
~ 14 pF/ft (42/m) |
Cable Temperature Range for V Probes, -98°F to 1600°F (-72°C to 870°C)
|
| Type T |
OD .063 (1.60) |
Semi-rigid Inconel 600 powder filled |
~ 60 pF/ft (180/m)
|
| Type N |
OD .142 (3.61) |
Flexible Inconel Braided |
~ 20 pF/ft (60/m)
|
| Type ND |
OD .165 (4.19) |
Low Impedance Shield Version of N |
~ 35 pF/ft (105/m) |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|