CYLINDRICAL SENSOR PROBE SERIES |
|
|
|
|
|
|
|
 |
|
|
|
HPC SERIES CYLINDRICAL PROBES
The cylindrical probes are the most popular version of the non-contact series. Its case is an additional component added to the sensor and guard elements to avoid being earth grounded. A set screw or split clamp housing is normally designed to hold it in place for easy adjustments. |
|
|
|
 |
|
|
|
|
STYLES
|
HPC-40
|
HPC-75
|
HPC-150
|
HPC-375
|
HPC-500
|
|
|
|
|
|
|
|
Sensor OD
|
0.040 (1.02)
|
0.075 (1.91)
|
0.150 (3.81)
|
0.375 (9.53)
|
0.500 (12.70)
|
|
Case OD
|
0.187 (4.75)
|
0.250 (6.35)
|
0.375 (9.53)
|
0.750 (19.05)
|
1.000 (25.40)
|
|
Linear Range
|
0.020 (0.51)
|
0.050 (1.27)
|
0.100 (2.54)
|
0.250 (6.35)
|
0.325 (8.26)
|
|
Max. Range
|
0.040 (1.02)
|
0.075 (1.91)
|
0.150 (3.81)
|
0.375 (9.53)
|
0.500 (12.70)
|
|
|
|
|
Specified in inches (millimeters)
Consult the Factory for complete specifications on calibrations and performance. |
|
|
|
 |
|
|
|
NOTES:
- The guard element is driven with an electronic voltage 2 to 20 Volts P - P maximum and must not be grounded.
- -D dip coatings also prevent the driver, sensor and guard elements from being accidentally earth grounded or shorted. (not available on "V" and "S" series probes)
- Range values tested with 5 ft. (1.524 meters) of Capacitec L2 cable.
- Range and signal to noise ratio will be degraded by longer cable lengths.
- A 20 point archived calibration record of voltage output versus nominal displacement is provided standard with every probe.
- Sensor must be calibrated with final length of cable for stated results.
- All components are electronically isolated from each other by greater than 20 megohms of leakage resistance (at room temperature) for best operation with the instrumentation.
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|